Abstract
Film coating of pharmaceutical dosage forms, such as tablets and pellets, can be used to tailor the drug release profile. With that regard, a uniform coating thickness of a single tablet (intra-tablet), all tablets (inter-tablet) and subsequent batches (inter-batch) is crucial. We present a method comparison between in-line (optical coherence tomography and near-infrared spectroscopy) and well-established off-line (height-, weight- and diameter-gain) approaches to determining the coating thickness of tablets. We used single tablets drawn during a commercial coating process. Comparing the low intra- and high inter-tablet coating variability indicated that the tablets had a broad distribution of spray zone passes but at a random tablet orientation. Even at the end of the coating process at a mean coating thickness of about 70 µm, the inter-tablet standard deviation was about 9 µm or 13% relative standard deviation. Determining correlations between the methods identified the factors that contribute to the measurement uncertainty and bias for each method. Ultimately, we aimed to establish that in-line methods match or even surpass the conventional off-line reference methods in terms of accuracy and precision of coating thickness measurement.
Original language | English |
---|---|
Pages (from-to) | 344-352 |
Number of pages | 9 |
Journal | European Journal of Pharmaceutics and Biopharmaceutics |
Volume | 142 |
DOIs | |
Publication status | Published - 1 Sep 2019 |
Keywords
- Coating variability
- Film coating
- Near-infrared spectroscopy (NIR)
- Optical coherence tomography (OCT)
- Process analytical technology (PAT)
ASJC Scopus subject areas
- Biotechnology
- Pharmaceutical Science