Simulated transcatheter aortic valve deformation: a parametric study on the impact of leaflet geometry on valve peak stress

Kewei Li, Wei Sun

Research output: Contribution to journalArticleResearchpeer-review

Abstract

In this study, we developed a computational framework to investigate the impact of leaflet geometry of a transcatheter aortic valve (TAV) on the leaflet stress distribution, aiming at optimizing TAV leaflet design to reduce its peak stress. Utilizing a generic TAV model developed previously [Li and Sun, Annals of Biomedical Engineering, 2010. 38(8): 2690–2701], we first parameterized the 2D leaflet geometry by mathematical equations, then by perturbing the parameters of the equations, we could automatically generate a new leaflet design, remesh the 2D leaflet model and build a 3D leaflet model from the 2D design via a Python script. Approximately 500 different leaflet designs were investigated by simulating TAV closure under the nominal circular deployment and physiological loading conditions. From the simulation results, we identified a new leaflet design that could reduce the previously reported valve peak stress by about 5%. The parametric analysis also revealed that increasing the free edge width had the highest overall impact on decreasing the peak stress. A similar computational analysis was further performed for a TAV deployed in an abnormal, asymmetric elliptical configuration. We found that a minimal free edge height of 0.46 mm should be adopted to prevent central backflow leakage. This increase of the free edge height resulted in an increase of the leaflet peak stress. Furthermore, the parametric study revealed a complex response surface for the impact of the leaflet geometric parameters on the peak stress, underscoring the importance of performing a numerical optimization to obtain the optimal TAV leaflet design.

Original languageEnglish
Article numbere02814
JournalInternational Journal for Numerical Methods in Biomedical Engineering
Volume33
Issue number3
DOIs
Publication statusPublished - 1 Mar 2017

Fingerprint

Aortic Valve
Geometry
Biomedical Engineering
Boidae
Parametric Analysis
Python
Biomedical engineering
Computational Analysis
Response Surface
Numerical Optimization
Solar System
Stress Distribution
Sun
3D Model
Leakage
Categorical or nominal
Design
Stress concentration
Closure
Configuration

Keywords

  • aortic stenosis
  • finite element analysis
  • heart valve biomechanics
  • optimization analysis
  • transcatheter aortic valve implantation

ASJC Scopus subject areas

  • Software
  • Modelling and Simulation
  • Biomedical Engineering
  • Molecular Biology
  • Computational Theory and Mathematics
  • Applied Mathematics

Cite this

Simulated transcatheter aortic valve deformation: a parametric study on the impact of leaflet geometry on valve peak stress. / Li, Kewei; Sun, Wei.

In: International Journal for Numerical Methods in Biomedical Engineering, Vol. 33, No. 3, e02814, 01.03.2017.

Research output: Contribution to journalArticleResearchpeer-review

@article{3148586dfec3490d841bab62fe68dd1d,
title = "Simulated transcatheter aortic valve deformation: a parametric study on the impact of leaflet geometry on valve peak stress",
abstract = "In this study, we developed a computational framework to investigate the impact of leaflet geometry of a transcatheter aortic valve (TAV) on the leaflet stress distribution, aiming at optimizing TAV leaflet design to reduce its peak stress. Utilizing a generic TAV model developed previously [Li and Sun, Annals of Biomedical Engineering, 2010. 38(8): 2690–2701], we first parameterized the 2D leaflet geometry by mathematical equations, then by perturbing the parameters of the equations, we could automatically generate a new leaflet design, remesh the 2D leaflet model and build a 3D leaflet model from the 2D design via a Python script. Approximately 500 different leaflet designs were investigated by simulating TAV closure under the nominal circular deployment and physiological loading conditions. From the simulation results, we identified a new leaflet design that could reduce the previously reported valve peak stress by about 5{\%}. The parametric analysis also revealed that increasing the free edge width had the highest overall impact on decreasing the peak stress. A similar computational analysis was further performed for a TAV deployed in an abnormal, asymmetric elliptical configuration. We found that a minimal free edge height of 0.46 mm should be adopted to prevent central backflow leakage. This increase of the free edge height resulted in an increase of the leaflet peak stress. Furthermore, the parametric study revealed a complex response surface for the impact of the leaflet geometric parameters on the peak stress, underscoring the importance of performing a numerical optimization to obtain the optimal TAV leaflet design.",
keywords = "aortic stenosis, finite element analysis, heart valve biomechanics, optimization analysis, transcatheter aortic valve implantation",
author = "Kewei Li and Wei Sun",
year = "2017",
month = "3",
day = "1",
doi = "10.1002/cnm.2814",
language = "English",
volume = "33",
journal = "International Journal for Numerical Methods in Biomedical Engineering",
issn = "2040-7939",
publisher = "Wiley-Blackwell",
number = "3",

}

TY - JOUR

T1 - Simulated transcatheter aortic valve deformation: a parametric study on the impact of leaflet geometry on valve peak stress

AU - Li, Kewei

AU - Sun, Wei

PY - 2017/3/1

Y1 - 2017/3/1

N2 - In this study, we developed a computational framework to investigate the impact of leaflet geometry of a transcatheter aortic valve (TAV) on the leaflet stress distribution, aiming at optimizing TAV leaflet design to reduce its peak stress. Utilizing a generic TAV model developed previously [Li and Sun, Annals of Biomedical Engineering, 2010. 38(8): 2690–2701], we first parameterized the 2D leaflet geometry by mathematical equations, then by perturbing the parameters of the equations, we could automatically generate a new leaflet design, remesh the 2D leaflet model and build a 3D leaflet model from the 2D design via a Python script. Approximately 500 different leaflet designs were investigated by simulating TAV closure under the nominal circular deployment and physiological loading conditions. From the simulation results, we identified a new leaflet design that could reduce the previously reported valve peak stress by about 5%. The parametric analysis also revealed that increasing the free edge width had the highest overall impact on decreasing the peak stress. A similar computational analysis was further performed for a TAV deployed in an abnormal, asymmetric elliptical configuration. We found that a minimal free edge height of 0.46 mm should be adopted to prevent central backflow leakage. This increase of the free edge height resulted in an increase of the leaflet peak stress. Furthermore, the parametric study revealed a complex response surface for the impact of the leaflet geometric parameters on the peak stress, underscoring the importance of performing a numerical optimization to obtain the optimal TAV leaflet design.

AB - In this study, we developed a computational framework to investigate the impact of leaflet geometry of a transcatheter aortic valve (TAV) on the leaflet stress distribution, aiming at optimizing TAV leaflet design to reduce its peak stress. Utilizing a generic TAV model developed previously [Li and Sun, Annals of Biomedical Engineering, 2010. 38(8): 2690–2701], we first parameterized the 2D leaflet geometry by mathematical equations, then by perturbing the parameters of the equations, we could automatically generate a new leaflet design, remesh the 2D leaflet model and build a 3D leaflet model from the 2D design via a Python script. Approximately 500 different leaflet designs were investigated by simulating TAV closure under the nominal circular deployment and physiological loading conditions. From the simulation results, we identified a new leaflet design that could reduce the previously reported valve peak stress by about 5%. The parametric analysis also revealed that increasing the free edge width had the highest overall impact on decreasing the peak stress. A similar computational analysis was further performed for a TAV deployed in an abnormal, asymmetric elliptical configuration. We found that a minimal free edge height of 0.46 mm should be adopted to prevent central backflow leakage. This increase of the free edge height resulted in an increase of the leaflet peak stress. Furthermore, the parametric study revealed a complex response surface for the impact of the leaflet geometric parameters on the peak stress, underscoring the importance of performing a numerical optimization to obtain the optimal TAV leaflet design.

KW - aortic stenosis

KW - finite element analysis

KW - heart valve biomechanics

KW - optimization analysis

KW - transcatheter aortic valve implantation

UR - http://www.scopus.com/inward/record.url?scp=84989244975&partnerID=8YFLogxK

U2 - 10.1002/cnm.2814

DO - 10.1002/cnm.2814

M3 - Article

VL - 33

JO - International Journal for Numerical Methods in Biomedical Engineering

JF - International Journal for Numerical Methods in Biomedical Engineering

SN - 2040-7939

IS - 3

M1 - e02814

ER -