Sensitivity maps and system requirements for magnetic induction tomography using a planar gradiometer

Javier Rosell, Roberto Casañas, Hermann Scharfetter

Research output: Contribution to journalArticleResearchpeer-review

Abstract

We evaluated analytically and experimentally the performance of a planar gradiometer as a sensing element in a system for magnetic induction tomography. A system using an excitation coil and a planar gradiometer was compared against a system with two coils. We constructed one excitation coil, two different sensing elements and a high-resolution phase detector. The first sensor was a PCB square spiral coil with seven turns. The second sensor was a PCB planar gradiometer with two opposite square spirals of seven turns, with a distance between centres of 8 cm. Theoretical sensitivity maps were derived from basic equations and compared with experimental data obtained at 150 kHz. The experimental sensitivity maps were obtained measuring the perturbation produced by a brass sphere of 12 mm in empty space. The advantage of using a gradiometer is that it can be adjusted to give a minimum signal for homogeneous objects, while increasing the sensitivity to local perturbations of the conductivity. Results show that a system using a planar gradiometer as detector has less demanding requirements for the electronic system than a system using simple coils.

Original languageEnglish
Pages (from-to)121-130
Number of pages10
JournalPhysiological measurement
Volume22
Issue number1
DOIs
Publication statusPublished - 14 Mar 2001

Fingerprint

Electromagnetic induction
Polychlorinated Biphenyls
Polychlorinated biphenyls
Tomography
Detectors
Sensors
Brass
brass

Keywords

  • Bioimpedance
  • Eddy currents
  • Electromagnetic tomography
  • Magnetic induction
  • Planar gradiometer

ASJC Scopus subject areas

  • Biophysics
  • Physiology
  • Biomedical Engineering
  • Physiology (medical)

Fields of Expertise

  • Human- & Biotechnology

Cite this

Sensitivity maps and system requirements for magnetic induction tomography using a planar gradiometer. / Rosell, Javier; Casañas, Roberto; Scharfetter, Hermann.

In: Physiological measurement, Vol. 22, No. 1, 14.03.2001, p. 121-130.

Research output: Contribution to journalArticleResearchpeer-review

@article{797d8ecf700241e094dfa5c2f85560ea,
title = "Sensitivity maps and system requirements for magnetic induction tomography using a planar gradiometer",
abstract = "We evaluated analytically and experimentally the performance of a planar gradiometer as a sensing element in a system for magnetic induction tomography. A system using an excitation coil and a planar gradiometer was compared against a system with two coils. We constructed one excitation coil, two different sensing elements and a high-resolution phase detector. The first sensor was a PCB square spiral coil with seven turns. The second sensor was a PCB planar gradiometer with two opposite square spirals of seven turns, with a distance between centres of 8 cm. Theoretical sensitivity maps were derived from basic equations and compared with experimental data obtained at 150 kHz. The experimental sensitivity maps were obtained measuring the perturbation produced by a brass sphere of 12 mm in empty space. The advantage of using a gradiometer is that it can be adjusted to give a minimum signal for homogeneous objects, while increasing the sensitivity to local perturbations of the conductivity. Results show that a system using a planar gradiometer as detector has less demanding requirements for the electronic system than a system using simple coils.",
keywords = "Bioimpedance, Eddy currents, Electromagnetic tomography, Magnetic induction, Planar gradiometer",
author = "Javier Rosell and Roberto Casa{\~n}as and Hermann Scharfetter",
year = "2001",
month = "3",
day = "14",
doi = "10.1088/0967-3334/22/1/316",
language = "English",
volume = "22",
pages = "121--130",
journal = "Physiological measurement",
issn = "0967-3334",
publisher = "IOP Publishing Ltd.",
number = "1",

}

TY - JOUR

T1 - Sensitivity maps and system requirements for magnetic induction tomography using a planar gradiometer

AU - Rosell, Javier

AU - Casañas, Roberto

AU - Scharfetter, Hermann

PY - 2001/3/14

Y1 - 2001/3/14

N2 - We evaluated analytically and experimentally the performance of a planar gradiometer as a sensing element in a system for magnetic induction tomography. A system using an excitation coil and a planar gradiometer was compared against a system with two coils. We constructed one excitation coil, two different sensing elements and a high-resolution phase detector. The first sensor was a PCB square spiral coil with seven turns. The second sensor was a PCB planar gradiometer with two opposite square spirals of seven turns, with a distance between centres of 8 cm. Theoretical sensitivity maps were derived from basic equations and compared with experimental data obtained at 150 kHz. The experimental sensitivity maps were obtained measuring the perturbation produced by a brass sphere of 12 mm in empty space. The advantage of using a gradiometer is that it can be adjusted to give a minimum signal for homogeneous objects, while increasing the sensitivity to local perturbations of the conductivity. Results show that a system using a planar gradiometer as detector has less demanding requirements for the electronic system than a system using simple coils.

AB - We evaluated analytically and experimentally the performance of a planar gradiometer as a sensing element in a system for magnetic induction tomography. A system using an excitation coil and a planar gradiometer was compared against a system with two coils. We constructed one excitation coil, two different sensing elements and a high-resolution phase detector. The first sensor was a PCB square spiral coil with seven turns. The second sensor was a PCB planar gradiometer with two opposite square spirals of seven turns, with a distance between centres of 8 cm. Theoretical sensitivity maps were derived from basic equations and compared with experimental data obtained at 150 kHz. The experimental sensitivity maps were obtained measuring the perturbation produced by a brass sphere of 12 mm in empty space. The advantage of using a gradiometer is that it can be adjusted to give a minimum signal for homogeneous objects, while increasing the sensitivity to local perturbations of the conductivity. Results show that a system using a planar gradiometer as detector has less demanding requirements for the electronic system than a system using simple coils.

KW - Bioimpedance

KW - Eddy currents

KW - Electromagnetic tomography

KW - Magnetic induction

KW - Planar gradiometer

UR - http://www.scopus.com/inward/record.url?scp=0035111772&partnerID=8YFLogxK

U2 - 10.1088/0967-3334/22/1/316

DO - 10.1088/0967-3334/22/1/316

M3 - Article

VL - 22

SP - 121

EP - 130

JO - Physiological measurement

JF - Physiological measurement

SN - 0967-3334

IS - 1

ER -