Solvent Resistant O2 Sensor Integrated in Pressured Flow Reactors

Philipp Sulzer, René Lebl, Torsten Mayr

Publikation: Beitrag in einer FachzeitschriftKonferenzartikelBegutachtung

Abstract

In this contribution we present an optical oxygen sensor that can easily be integrated into continuous flow reactors. The sensors are pressure resistant to up to 8 bars and are able to continuously measure oxygen for several hours in organic solvents. The dynamic range was evaluated up to 42 mmol/L O2 and several calibrations showed the sensors to perform linearly according to Stern-Volmer, which allows convenient two-point calibrations. By using state-of-the-art HPLC connectors and materials the sensor can be implemented at virtually every point of a standard flow reactor. Readout is conducted via a 4-channel USB phase fluorimeter.
Originalspracheenglisch
Aufsatznummer994
Seitenumfang4
FachzeitschriftProceedings MDPI AG
Jahrgang2
Ausgabenummer13
DOIs
PublikationsstatusVeröffentlicht - 16 Jan. 2019
VeranstaltungEurosensors 2018 - Karl Franzens Universität Graz, Graz, Österreich
Dauer: 9 Sept. 201812 Sept. 2018
https://www.eurosensors2018.eu

Fingerprint

Untersuchen Sie die Forschungsthemen von „Solvent Resistant O2 Sensor Integrated in Pressured Flow Reactors“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren