Optimal parameters in variable-velocity scanning luminescence lifetime microscopy

Zdenek Petrasek*, Juan M. Bolivar, Bernd Nidetzky*

*Korrespondierende/r Autor/-in für diese Arbeit

Publikation: Beitrag in einer FachzeitschriftArtikelBegutachtung

Abstract

We determine the optimal parameters (scan velocities) for measuring the luminescence lifetime on the microsecond scale using the recently introduced method based on scanning the excitation beam across the sample. Using simulations, we evaluate the standard deviation and bias of the luminescence decay rate determined by scanning with two different velocities. The analysis is performed for Poisson- and normal-distributed signals, representing different types of detection techniques. We also show that a weak uncorrected background induces a bias in the obtained decay rate, and take this effect into account when choosing optimal measurement parameters. For comparison, the analysis is additionally performed for two conventional gating schemes for lifetime measurement. The variable-velocity scanning method is found to be more robust to the effect of the background signal than the gating schemes.

Originalspracheenglisch
Seiten (von - bis)71-78
Seitenumfang8
FachzeitschriftMicroscopy Research and Technique
Jahrgang84
Ausgabenummer1
DOIs
PublikationsstatusVeröffentlicht - Jan. 2021

ASJC Scopus subject areas

  • Instrumentierung
  • Medizinische Labortechnik
  • Anatomie
  • Histologie

Fingerprint

Untersuchen Sie die Forschungsthemen von „Optimal parameters in variable-velocity scanning luminescence lifetime microscopy“. Zusammen bilden sie einen einzigartigen Fingerprint.