Optical Ammonia Sensor for Continuous Bioprocess Monitoring

Maximilian Maierhofer, Sergey Borisov, Torsten Mayr

Publikation: Beitrag in Buch/Bericht/KonferenzbandBeitrag in einem KonferenzbandBegutachtung

Abstract

We present an optical ammonia sensor suitable for bioprocess monitoring. A fluorescent dye is physically entrapped in a polyurethane hydrogel forming an emulsion system with vinylterminated polydimethylsiloxane (PDMS). The sensing layer is covered by a hydrophobic porous membrane which excludes hydrophilic substances. Ammonia, diffuses through this barrier and PDMS to the protonated dye, whereby it deprotonates the dye and switches off its emission. Readout is performed with a miniaturized phase fluorimeter combined with optical fibers. Dual-lifetime referencing (DLR) acts as detection method and Egyptian blue as reference material.
Originalspracheenglisch
TitelProceedings Eurosensors 2018
Redakteure/-innenAnton Köck, Marco Deluca
Herausgeber (Verlag)MDPI AG
Seitenumfang4
Band2
PublikationsstatusVeröffentlicht - 19 Nov. 2018
VeranstaltungEurosensors 2018 - Karl Franzens Universität Graz, Graz, Österreich
Dauer: 9 Sept. 201812 Sept. 2018
https://www.eurosensors2018.eu

Konferenz

KonferenzEurosensors 2018
Land/GebietÖsterreich
OrtGraz
Zeitraum9/09/1812/09/18
Internetadresse

Schlagwörter

  • Optischer Sensor
  • Ammoniak
  • Bioprozessüberwachung
  • Fluoreszenz

Fingerprint

Untersuchen Sie die Forschungsthemen von „Optical Ammonia Sensor for Continuous Bioprocess Monitoring“. Zusammen bilden sie einen einzigartigen Fingerprint.

Dieses zitieren