In four-dimensional time-dependent computed tomography
(4D-CT) of the lungs, irregularities in breathing
movements can cause errors in data acquisition, or even
data loss. We present a method based on sending a synthetic,
regular breathing signal to the CT instead of the
real signal, which ensures 4D-CT data sets without data
loss. Subsequent correction of the signal based on the real
breathing curve enables an accurate reconstruction of the
size and movement of the target volume. This makes it possible
to plan radiation treatment based on the obtained
data. The method was tested with dynamic thorax phantom
measurements using synthetic and real breathing patterns:
Titel in ÜbersetzungMethodik zur verbesserten Datenerfassung bei der 4D-Computertomographie für die stereotaktische ablative Strahlentherapie
Originalspracheenglisch
Seiten (von - bis)31-38
FachzeitschriftZeitschrift für medizinische Physik
Jahrgang27
Ausgabenummer1
DOIs
PublikationsstatusVeröffentlicht - 2017

Schlagwörter

  • Unregelmäßige Atmung
  • Datenverlustfrei
  • 4D-CT
  • SABR
  • 4D-CT Rekonstruktionen
  • Atmungserkennung

Dies zitieren

A method for improved 4D-computed tomography data acquisition. / Kupper, Martin; Sprengel, Wolfgang; Winkler, Peter; Zurl, Brigitte.

in: Zeitschrift für medizinische Physik, Jahrgang 27, Nr. 1, 2017, S. 31-38.

Publikation: Beitrag in einer FachzeitschriftArtikelForschungBegutachtung

@article{4f3630b07ae648dd87a483be3d76c472,
title = "A method for improved 4D-computed tomography data acquisition",
abstract = "In four-dimensional time-dependent computed tomography(4D-CT) of the lungs, irregularities in breathingmovements can cause errors in data acquisition, or evendata loss. We present a method based on sending a synthetic,regular breathing signal to the CT instead of thereal signal, which ensures 4D-CT data sets without dataloss. Subsequent correction of the signal based on the realbreathing curve enables an accurate reconstruction of thesize and movement of the target volume. This makes it possibleto plan radiation treatment based on the obtaineddata. The method was tested with dynamic thorax phantommeasurements using synthetic and real breathing patterns:",
keywords = "Unregelm{\"a}{\ss}ige Atmung, Datenverlustfrei, 4D-CT, SABR, 4D-CT Rekonstruktionen, Atmungserkennung",
author = "Martin Kupper and Wolfgang Sprengel and Peter Winkler and Brigitte Zurl",
year = "2017",
doi = "10.1016/j.zemedi.2016.05.001",
language = "English",
volume = "27",
pages = "31--38",
journal = "Zeitschrift f{\"u}r medizinische Physik",
issn = "0939-3889",
publisher = "Urban und Fischer Verlag Jena",
number = "1",

}

TY - JOUR

T1 - A method for improved 4D-computed tomography data acquisition

AU - Kupper, Martin

AU - Sprengel, Wolfgang

AU - Winkler, Peter

AU - Zurl, Brigitte

PY - 2017

Y1 - 2017

N2 - In four-dimensional time-dependent computed tomography(4D-CT) of the lungs, irregularities in breathingmovements can cause errors in data acquisition, or evendata loss. We present a method based on sending a synthetic,regular breathing signal to the CT instead of thereal signal, which ensures 4D-CT data sets without dataloss. Subsequent correction of the signal based on the realbreathing curve enables an accurate reconstruction of thesize and movement of the target volume. This makes it possibleto plan radiation treatment based on the obtaineddata. The method was tested with dynamic thorax phantommeasurements using synthetic and real breathing patterns:

AB - In four-dimensional time-dependent computed tomography(4D-CT) of the lungs, irregularities in breathingmovements can cause errors in data acquisition, or evendata loss. We present a method based on sending a synthetic,regular breathing signal to the CT instead of thereal signal, which ensures 4D-CT data sets without dataloss. Subsequent correction of the signal based on the realbreathing curve enables an accurate reconstruction of thesize and movement of the target volume. This makes it possibleto plan radiation treatment based on the obtaineddata. The method was tested with dynamic thorax phantommeasurements using synthetic and real breathing patterns:

KW - Unregelmäßige Atmung

KW - Datenverlustfrei

KW - 4D-CT

KW - SABR

KW - 4D-CT Rekonstruktionen

KW - Atmungserkennung

UR - http://www.sciencedirect.com/science/article/pii/S0939388916300290

U2 - 10.1016/j.zemedi.2016.05.001

DO - 10.1016/j.zemedi.2016.05.001

M3 - Article

VL - 27

SP - 31

EP - 38

JO - Zeitschrift für medizinische Physik

JF - Zeitschrift für medizinische Physik

SN - 0939-3889

IS - 1

ER -